Technical appendix

A Vision Opportunity: Technical Appendix

1

Summary

A head-mounted display is an image-plane corrector; it can apply a different geometric correction at every point of the visual field, which spectacles cannot. That makes Apple Vision Pro a plausible platform for the epiretinal-membrane (ERM) distortion, the genuinely field-dependent part of Patrick’s condition. Intermittent exotropia is a different kind of problem: a whole-field misalignment that switches on and off. It needs fast, reliable per-eye tracking more than field-dependent correction.

The decisive engineering constraint is gaze registration. ERM distortion is fixed to the retina, so the correction must move with the eye, and its accuracy is limited by eye-tracking error relative to the distortion’s spatial scale (see the error budget in section 4). Because Apple Vision Pro does not expose gaze data to developers [1], only Apple can build this at the system level.

2

Primer: pupil-plane vs. image-plane correction

Conventional eyewear works because most of the eye’s refractive error is isoplanatic: the same wavefront error applies, to good approximation, at every field angle. A single lens near the pupil can then correct the whole field at once.

A spectacle lens can therefore remove terms that are common across the field:

  • defocus and astigmatism (wavefront terms)
  • a uniform angular shift (prism)
  • a uniform or meridional magnification (iseikonic lenses, for aniseikonia)

It cannot apply an arbitrary field-dependent displacement map d(θ), where θ is visual direction. That would require the correction to differ at every field point, which is the definition of an anisoplanatic problem.

A display behaves the opposite way. Rendering an image is a mapping from scene direction to display pixel, so a display can apply any warp

Ishown(θ) = Iscene (θ+d(θ))

with d(θ) free to vary point by point. What a display cannot do is correct blur within each point’s image. It emits incoherent light ahead of the eye’s own optics, so it controls where light lands, not the wavefront inside each point spread function.

The two correctors are complementary: lenses handle isoplanatic wavefront error, and displays handle field-dependent geometry. Apple Vision Pro already pairs them, with ZEISS prescription inserts for refraction and a fully programmable image path.

3

The two conditions in this framework

Neither condition is an optical aberration in the usual sense, and they sit in different places on the isoplanatic map.

Intermittent exotropiaEpiretinal membrane
Physical originOculomotor: one eye’s pointing direction drifts outwardRetinal: traction displaces the sampling layer tangentially
Optical analogueTime-varying whole-field tilt (a fast-steering-mirror problem)Warped detector pixel map (a geometric-calibration problem)
Field dependenceApproximately uniform; may vary with gaze directionStrongly field-dependent across the central field
Time dependenceIntermittent, switching on and offSlowly varying (months to years)
Coordinate frameHead/scene frame, applied to one eyeRetinal frame, so it moves with every eye movement
What lenses manageStatic prism only, a fixed compromiseUniform magnification at best
What a display addsA prism that switches on when drift is detectedAn arbitrary inverse warp for the affected eye
Key requirementReliable per-eye tracking of the deviating eyeGaze registration fine compared with the distortion scale

For ERM, the leading explanations involve tangential displacement of photoreceptors or Müller cells, with damage to the inner nuclear layer also implicated; the full mechanism is not settled [2]. Either way, the defect is in where the retina samples the image, not in the light reaching it. That is why it is better framed as detector-plane calibration than as wavefront correction or imaging through random media.

For exotropia, the correction is a prism that tracks the deviation in time. The technology is simpler than for ERM, but the clinical question is harder (see the open questions in section 6).

4

Error budget for gaze-contingent distortion correction

The ERM correction helps only if gaze error is small compared with the distortion’s characteristic length scale. Because the distortion d(ρ) is defined in retinal coordinates ρ, the display must place the inverse warp at the current gaze direction. A gaze-tracking error δ misregisters the correction, leaving a residual distortion of approximately

r(ρ) ≈ d(ρ+δ) − d(ρ) ≈ ∇d(ρ)δ

The residual depends on the gradient of the distortion map, not its amplitude. Correction improves on no correction only where |r| < |d|, which defines a requirement on the tracker:

|δ| ≪ L , L= |d| |∇d|

Here L is the local spatial scale over which the distortion changes. Smooth, broad distortions (large L) tolerate tracking error. Sharp folds near the fovea (small L) do not, and a misregistered correction there could add distortion rather than remove it.

Independent measurements put Apple Vision Pro’s gaze accuracy at about 0.45° to 1.1° in central viewing, worsening to several degrees in some tasks [3], [4]. Those figures are averages over normal eyes and include calibration bias. What matters here is precision and latency for one eye of one user, against an L measured from Patrick’s own distortion map. Fixational eye movements of a fraction of a degree add to δ unless the tracker follows them.

Two measurements would settle feasibility before any build:

  1. Patrick’s distortion map d(ρ), from subjective mapping (Amsler or M-CHARTS-style tasks in the headset) cross-checked against en-face OCT displacement, giving L across the central field.
  2. The per-eye gaze precision and latency that Apple’s tracker achieves internally on his eyes, which only Apple can measure.
5

Platform constraints on Apple Vision Pro today

The hardware is close to what this needs; the software access is not, which is why this has to be an Apple-level effort.

FactorStatusImplication
Passthrough latencyAbout 11 ms photon-to-photon, against 35–40 ms for competing headsets [5]Fast enough to warp the live scene without obvious lag
Gaze data accessNot exposed to apps; a narrow entitlement gives only approximate gaze for foveated streaming [1]A third party cannot build gaze-contingent correction; Apple can
Camera accessEnterprise APIs open the main cameras, not gaze [6]A static (non-gaze-contingent) warp is buildable now as a research prototype
Strabismus supportApple notes eye misalignment can impair eye detection and offers single-eye input [7]Apple already recognises the case; correction needs tracking of the deviating eye, not just the fixing eye
Focal distanceFixed accommodation distance for all contentVergence and accommodation are decoupled, which interacts with exotropia control

The enterprise camera route matters for a first study. A static warp, re-centred by the user’s fixation on a target, could test whether Patrick perceives a reduction in distortion before Apple commits engineering effort to gaze-contingent rendering.

6

Prior art and open clinical questions

Both ideas have been prototyped elsewhere; what Apple Vision Pro could add is the latency, display quality and system-level control to make them practical.

WorkWhat it doesRelevance
MASCARA, Newcastle University (UK) [8]Users define their own warping map; its inverse is applied to the affected eye in real time in a commercial headsetClosest precedent for ERM correction
Ling et al., University of Surrey [9]Toolkit of monocular and binocular methods for mapping and correcting metamorphopsiaDistortion-mapping methods
Aegenta VR trial [10]Headset-based measurement of metamorphopsia and aniseikonia, including ERM patientsMeasurement protocol
Norcia, Stanford Bio-X [11]Gaze-contingent VR keeping each eye’s image on corresponding retinal points in strabismusClosest precedent for vergence correction
STARE, Moorfields / City, University of London [12]Automated strabismus measurement with an eye-tracked AR headsetDeviation measurement

Open clinical questions a partner would need to answer:

  • Fusion vs. dependence. If the image follows the drifting eye, the disparity signal that drives fusional recovery disappears. This could stabilise the deviation rather than reduce it, so a tapered or therapeutic mode may be needed.
  • Cortical adaptation. The brain partly adapts to long-standing metamorphopsia, so an exact optical inverse may not be the perceptually correct one.
  • Binocular interaction. Correcting one eye’s distortion changes what each eye contributes to fusion, which interacts with the exotropia.
  • Accommodation. A fixed focal distance removes the accommodative cues that often help control intermittent exotropia.
7

Proposed team and first study

The team covers wavefront sensing, correction physics, holographic optics and ophthalmic optics. The gap to fill before approaching Apple formally is a clinical partner.

RoleWhoContribution
Research subject and systems engineerPatrick BurrellDetailed, repeatable reports of perceived distortion and diplopia
Wavefront sensing and correction physicsDr. Derek Burrell, Fraunhofer IOSBWavefront sensing, field-dependent correction and error budgets; computational imaging and physics-informed deep learning for distortion mapping and registration
Holographic optics, lens design and ocular opticsDr. Kevin Murphy, TU DublinHolographic optical elements, lens design, vision correction and ophthalmic optics
Clinical lead (to recruit)Retina specialist and orthoptistDistortion mapping, safety, and the fusion questions above
PlatformApple Vision Pro teamGaze access, system-level rendering, engineering support

A realistic first study is single-subject and low-risk:

  1. Map Patrick’s distortion in each eye with headset-based subjective tests, cross-checked against OCT.
  2. Compute L across the central field and compare it with the tracker’s measured error, giving a go/no-go on gaze-contingent correction.
  3. Test a static inverse warp on a fixation target and score perceived distortion before and after.
  4. Only then, with clinical oversight, trial time-varying prism for the exotropia.

The MASCARA group at Newcastle has already built the subjective-mapping step and is the natural first clinical contact.

8

References

  1. [1]Apple Developer Forums, “Opening gaze data to streamer developers?” [Online]. Available: https://developer.apple.com/forums/thread/833150. Accessed: Oct. 20, 2026.
  2. [2]R. Matoba and Y. Morizane, “Epiretinal membrane: an overview and update,” Jpn. J. Ophthalmol., vol. 68, no. 6, pp. 603–613, Nov. 2024, doi: 10.1007/s10384-024-01127-6.
  3. [3]Z. Huang, G. Zhu, X. Duan, Y. Li, R. Wang, S. Zhang, and Z. Wang, “The eye-tracking accuracy of Apple Vision Pro and its impact on perceived usability,” Virtual Reality, Jul. 2026, doi: 10.1007/s10055-026-01442-y.
  4. [4]E. Tomiuk, J. Miró, and L. Duong, “Evaluation of eye-tracking capabilities in Apple Vision Pro for training in hybrid ventricular septal defect procedures,” in Proc. SPIE Medical Imaging 2025: Image-Guided Procedures, Robotic Interventions, and Modeling, M. E. Rettmann and J. H. Siewerdsen, Eds., 2025, doi: 10.1117/12.3047251.
  5. [5]9to5Mac, “Vision Pro latency the best on one metric; lags behind Meta on another,” Feb. 16, 2024. [Online]. Available: https://9to5mac.com/2024/02/16/vision-pro-latency/
  6. [6]UploadVR, “Apple will let enterprise companies access Vision Pro’s passthrough cameras.” [Online]. Available: https://uploadvr.com/apple-vision-pro-enterprise-raw-camera-access. Accessed: Oct. 20, 2026.
  7. [7]Apple Support, “Using Apple Vision Pro with vision prescriptions and vision conditions.” [Online]. Available: https://support.apple.com/en-us/HT213965. Accessed: Oct. 20, 2026.
  8. [8]NHS Health Research Authority, “MetAmorphopSia Correction via Augmented ReAlity (MASCARA), Version 1.0,” IRAS ID 314997, Newcastle University. [Online]. Available: https://www.hra.nhs.uk/…/metamorphopsia-correction-via-augmented-reality-mascara-version-10/
  9. [9]Y. Ling, D. M. Frohlich, T. H. Williamson, and J.-Y. Guillemaut, “A toolkit of approaches for digital mapping and correction of visual distortion,” in Proc. 25th Int. ACM SIGACCESS Conf. Computers and Accessibility (ASSETS ’23), New York, NY, USA, 2023, Art. no. 60, doi: 10.1145/3597638.3614510.
  10. [10]ClinicalTrials.gov, “Virtual Reality Aniseikonia and Metamorphopsia Testing,” NCT07838181. [Online]. Available: https://clinicaltrials.gov/study/NCT07838181
  11. [11]Stanford Bio-X, “Interdisciplinary Initiatives Program Round 9 – 2018” (A. Norcia). [Online]. Available: https://biox.stanford.edu/research/seed-grants/…
  12. [12]N. Nixon, P. B. M. Thomas, and P. R. Jones, “Feasibility study of an automated strabismus screening test using augmented reality and eye-tracking (STARE),” Eye, vol. 37, no. 17, pp. 3609–3614, Dec. 2023, doi: 10.1038/s41433-023-02566-0.